Testing the Validity of the Averaged Approximation for the IAsys

Author(s)Edwards, David A.
Author(s)Jackson, S.A.
Date Accessioned2005-02-16T20:18:48Z
Date Available2005-02-16T20:18:48Z
Publication Date2002
AbstractOne device used to measure rate constants is the IAsys, and the flow in such a device can be modeled as stagnation point flow. Due to the special nature of the flow, the effects of transport on a surface reaction near a stagnation point may be incorporated exactly as long as the initial concentration of bound state is uniform. However, if the bound state is nonuniform initially, a complicated integrodifferential equation arises for the evolution of the bound state. Such a form is inconvenient for data analysis. The averaged approximation replaces the nonuniform initial state with its average, thus simplifying the analysis. This approximation is correct to O(Da) as the Damköhler number Da —> 0. A numerical simulation of the integrodifferential equation is performed which shows that the averaged approximation is useful even outside this regime.en
SponsorThis work was supported by the University of Delaware Research Foundation.en
Extent170554 bytes
MIME typeapplication/pdf
URLhttp://udspace.udel.edu/handle/19716/327
Languageen_US
PublisherDepartment of Mathematical Sciencesen
Part of SeriesTechnical Report: 2002-03
KeywordsSurface-volume reactionsen
Keywordsintegrodifferential equationsen
Keywordsnumerical simulationen
Keywordssurface plasmon resonanceen
KeywordsIAsysen
TitleTesting the Validity of the Averaged Approximation for the IAsysen
TypeTechnical Reporten
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
tcrpt_2002_03.pdf
Size:
166.56 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
2.31 KB
Format:
Item-specific license agreed upon to submission
Description: